HTS - Important Considerations
Assay type
- Biochemical assays
- Cell-based assays (2D, 3D spheroids, organoids)
Readout/measurement type
- High-content (microscopy)
- Spectroscopic (fluorescence, luminescence, absorbance)
Plate format
- Plate type 96- or 384-well format
- Plate color: transparent, white, black
- Well shape: flat, round, V-shaped
- Plate material: polystyrene, polypropylene, glass, etc.
- Coating: none, TC, ultra-low attachment, etc.
Controls
- Selection of appropriate negative- and positive controls
- Solvent concentration (e.g. 0.1 % DMSO, max. 0.3 % for cell-based assays)
- Number of controls and replicates
Assay Quality
- Z’-factor: measure of assay quality based on the lower and upper limit of the assay and the standard deviations of the positive and negative controls.
- Z’ value 0.5-1.0: excellent assay
- Z’ value 0.0 - 0.5: poor to acceptable assay
- Z’ value < 0.0: assay not working since positive and negative controls overlap
Handling capabilities of the equipment
- Minimum volumes
- Dead volumes
- Cooling/heating of media or reagents
- Edge effects
Assay Data Analysis
- Correction of edge effects and cell seeding patterns
- Normalization
- Hit selection
Data storage
- Imaging-based assays require large storage capacity compared to spreadsheet-based data